The Resmon Pro is a next-generation respiratory assessment system designed to accurately and non-invasively measure respiratory mechanics using the Forced Oscillation Technique, or Oscillometry. Built for modern clinical environments, the Resmon Pro provides fast, simple, and effort-independent testing with only just a few normal tidal breaths.
The Resmon Pro is suitable for adults and children aged three years and older.
Key Features:
Effort-Independent Respiratory Assessment
- No forced manoeuvres required.
- Detection and localisation of airway obstruction
- Differentiation between central, peripheral, and heterogeneous types of obstruction.
Dynamic "Within-Breath" Testing
- Real-time visualisation of Resistance (Rrs), Reactance (Xrs), and flow or volume (configurable)
- Measurement of inspiratory, expiratory, and total Rrs and Xrs parameters.
- Automatic detection of tidal expiratory flow limitation with percentage of flow limited breaths
Slow Vital Capacity (SVC) & Lung Volume Insights
- SVC for monitoring of restrictive patterns
- Inspiratory Capacity for hyperinflation detection
- Closing Volume (CVfot) and Critical Reactance (Xcrit) for evaluating de-recruitments of peripheral airways
- Specific airway conductance (sGrs) for the evaluation of therapeutic intervention compensated for changes in lung volumes
Multiple Testing Modes
- Multi-frequency mode: 5-11-19 Hz
- Single-frequency mode: 5, 6, 8, 10 Hz
- Pseudo Random Noise: 5-37 Hz
Enhanced Patient Comfort & Testing Efficiency
- Automatic adjustment of stimulus amplitude based on patient impedance
- Optimises measurement quality and patient comfort
- Automatic discarding of breaths affected by artifacts
- User-selectable number of accepted breaths, or time to end measurement automatically to minimise operator intervention.
- Coefficient of variability (CoV%) within-measurement and within-session for quality control.
- Selectable reference equations for predicted normal values and Z-score calculation
RESTECH: Resmon Pro
Flow Sensor:
- Range: ± 1.5 L/s
Metal Mesh:
- Linearity: < ±2% within range ±1.5 L/s
Pressure:
- Range: ±2.5 kPa
- Linearity: 1.5 %fs
Dimensions & Weight:
- 31 cm x 29 cm x 26 cm
- 6.4 kg with support arm; 4.3 kg device only
Testing Signal mode:
- Single Frequency: 5, 6, 8, 10 Hz
- Enhanced & Optimised Multi-Frequency: 5-11-19 Hz
- Pseudo Random Noise: 5-37 Hz
Measurement Accuracy:
- Rrs, Xrs: ≤ ±0.1 cmH2O/L/s or ≤ ±10% of the measured value
- Breathing pattern parameters: ≤ ±10% of the measured value
- Lung Volumes: ≤ ±50 mL or ≤ ±2.5% of the measured value
- Environmental Conditions:
- Temperature: ±1°C
- Humidity: ±3% (relative humidity)
- Pressure: ±100 Pa
Calibration:
- Factory calibration according to international recommendations ERS (Eur Respir J. 2020;55(2):1900753). Daily calibration check with a test object (supplied with the device) and 3 liters calibration syringe (optional not supplied with the device), required for the measurement of slow spirometry volumes, closing volumes and airways specific conductance.
Patient Load:
- 0.68 - 0.95 cmH2O/L/s in the frequencies of normal breathing (0.2 - 0.98 Hz)
Dead Space:
- 35 mL
Display:
- 10.1" HD colour display with multi-touch capacitive touchscreen and anti-glare film (can be used with medical gloves)
Electrical Specifications:
- Medical grade 100/240 V, 50/60 Hz 60 W input AC/15 VDC output power supply (included)